AccuLine 40-6926 Руководство по эксплуатации - Страница 6
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Лазерный уровень AccuLine 40-6926. AccuLine 40-6926 12 страниц. 26x automatic level
1316H-English
7/10/08
9:45 AM
Page 6
5. Calibration
5.1 Compensator Check Button
Check the compensator for proper operation before use or anytime
the operation of the instrument is in question. Push and release the
compensator check button to shake the compensator. The compen-
sator should return to the exact horizontal position sighted before the
check button was pressed.
5.2 Circular Bubble Vial
Center the vial bubble using the leveling screws, then rotate the
instrument 180º. The bubble should remain centered (Fig. 3). If the
bubble moved from the center, the vial needs adjustment (Fig. 4).
Fig. 3
Fig. 4
Turn the leveling screws to bring the bubble halfway to center (Fig. 5).
Using the allen wrench, turn the two vial adjustment screws to center
the bubble (Fig. 6).
Repeat the above procedure until the bubble remain centered when
the level is rotated 180º.
Fig. 6
Fig. 5
6
©2008 Johnson Level & Tool