Olympus BH-2 BHTU Руководство - Страница 5

Просмотреть онлайн или скачать pdf Руководство для Микроскоп Olympus BH-2 BHTU. Olympus BH-2 BHTU 24 страницы. Sliding focus block
Также для Olympus BH-2 BHTU: Полная разборка, очистка и сборка (23 страниц)

Olympus BH-2 BHTU Руководство
 
Filtered DC Power Supply 
Diode D201 and capacitor C201 operate as a peak detector to produce filtered DC power from the full‐wave‐rectified 
output  of  bridge  rectifier  DB101.    This  filtered  DC  power  line  supplies  power  to  the  comparators  and  their  respective 
LEDs.    The  AC  ripple  present  on  this  line  depends  on  the  amount  of  current  pulled  by  the  comparators  and  LEDs.    At 
minimal commanded lamp intensity (i.e., with all four LEDs in the off state), very little current is required and the voltage 
on  this  line  is  approximately  +10.5V  with  little  AC  ripple.    As  the  lamp  intensity  is  increased  (and  the  LEDs  begin  to 
successively  illuminate),  the  current  drain  on  this  line  increases  as  does  the  resulting  AC  ripple.    When  the  lamp  is  at 
maximum intensity, all four LEDs are illuminated and the AC ripple is at its maximum, resulting in an average voltage on 
the filtered DC power supply line of approximately +9.5V.  Figure 25 shows the ripple voltage on the filtered DC power 
supply  line  with  the  equipment  operating  under  worst‐case  ripple  conditions  of  maximum  lamp  intensity.    The  green 
trace is the positive output of bridge rectifier DB101 and the blue trace is the filtered DC power supply line. 
Comparator Thresholds 
The  switching  thresholds  of  the  four  op‐amp  comparators  are  provided  by  a  four‐stage  voltage  divider  consisting  of 
resistors R208, R209, R210, R211, and R212.  These thresholds are configured to allow the comparators to illuminate the 
LEDs at pre‐defined lamp voltages of 2V, 4V, 6V, and 7V, in response to the decreasing control signal from the intensity 
potentiometer.  Figure 26 shows the threshold voltages on the  inverting  pins of the op‐amp comparators.  The green 
trace is the positive output of bridge rectifier DB101, from which the four threshold voltages are derived.  The blue, red, 
cyan,  and  magenta  traces  are  the  threshold  voltages  on  the  inverting  inputs  of  IC1‐A,  IC1‐D,  IC1‐B,  and  IC1‐C, 
respectively. 
Control‐Signal Clamp 
Resistors R213 and R214, and diodes D202, D203, and D204, clamp the voltage level of the intensity control signal from 
potentiometer  RV101  to  keep  the  non‐inverting  inputs  of  the  comparators  from  ever  dropping  below  approximately 
+0.7V.    This  is  necessary  to  prevent  the  LEDs  from  illuminating  briefly  during  the  intervals  where  both  the  wiper  of 
potentiometer RV101 and the comparator thresholds drop to zero at the zero crossings of the AC line.  If both inputs to 
the comparators were allowed to simultaneously drop to zero, the comparators would be in an indeterminate state, and 
depending  on  the  input‐offset  voltage  parameters  of  the  specific  op‐amp  devices,  their  outputs  could  pull  low  during 
these intervals, resulting in one or more LEDs not fully extinguishing when they should. 
The  operation  of  the  control‐signal  clamp  is  as  follows.    Resistor  R214  feeds  a  DC  current  through  the  two  series‐
connected  silicon  diodes,  D203  and  D204,  producing  a  potential  on  the  anode  of  diode  D204  two  diode  drops  above 
ground (approximately +1.4V).  The anode of diode D202 connects to this +1.4V node and its cathode connects to the 
non‐inverting inputs of the four comparators.  Resistor R213 couples the control signal from potentiometer RV101 to the 
non‐inverting  inputs  of  the  comparators.    Any  time  the  control  signal  is  low  enough  to  forward  bias  diode  D202,  the 
voltage on the non‐inverting inputs of the comparators is clamped to approximately +0.7V (i.e., one diode drop less than 
+1.4V).  As the voltage on the wiper of RV101 increases, diode D202 turns off and is effectively out of the circuit, thereby 
providing  no  further  clamping  action.    This  mechanism  allows  the  intensity  control  signal  to  swing  positive  without 
restraint,  but  prevents  it  from  dropping  below  approximately  +0.7V.   
Figure 
27  shows  the  waveform  of  the  clamped 
control signal at the 50% setting of the intensity control, as measured at the non‐inverting inputs of the comparators. 
Voltage Comparators 
Integrated  circuit  IC1  is  a  NEC  uPC324C  quad  operation  amplifier.    All  four  sections  of  IC1  are  operated  open‐loop  as 
voltage comparators and are configured to drive the four bar‐graph LEDs in response to the clamped control signal from 
potentiometer  RV101.    Each  comparator  illuminates  its  associated  LED  by  driving  its  output  low  during  the  intervals 
where the clamped control signal drops below its threshold voltage, and extinguishes its LED by driving its output high 
during the intervals where the clamped control signal exceeds its threshold voltage.  Figure 28 shows the operation of 
one of the comparators (IC1‐B) with the 6V LED illuminated.  In these waveforms, the intensity control is set to the 70% 
setting.  The green waveform is the threshold voltage present on the inverting pin of IC1‐B, and the blue waveform is the 
clamped  control  signal  on  the  non‐inverting  pin  of  IC1‐B.    During  the  intervals  where  the  control  signal  is  below  the 
threshold voltage, the comparator output pulls low, illuminating the 6V LED, and during the intervals where the control 
signal  is  clamped  to  approximately  +0.7V  (and  is  exceeding  the  threshold  voltage),  the  comparator  output  pulls  high, 
Olympus BH‐2 (BHT/BHTU) Electronics 
 
 
   Revision 2 
   
 
 
        Page 5 of 24