Olympus BH2-5RE Полная разборка, очистка и сборка - Страница 13
Просмотреть онлайн или скачать pdf Полная разборка, очистка и сборка для Микроскоп Olympus BH2-5RE. Olympus BH2-5RE 18 страниц. Modular revolving nosepiece
Reinstall the Slotted Lock Ring
Carefully engage the threads of the slotted lock ring
onto the pivot-adjustment screw (see
Figure 33 – Reinstall the slotted lock ring
Spin the revolving turret until one of the five threaded
objective bores in the revolving turret aligns with the
bore in the stationary base. Insert a suitably sized
screwdriver handle into these bores to lock the
revolving turret to the stationary base. Use a suitable
tool (see
Figure
9) to snug the slotted lock ring down to
lock the pivot-adjustment screw in place (see
34). Do not allow the pivot-adjustment screw to rotate
while tightening the slotted lock ring.
Figure 34 – Tighten the slotted lock ring
Verify the Feel of the Revolving Nosepiece
Hold the turret assembly by gripping the stationary base
in one hand and spin the revolving turret with the other
hand. The motion of the turret should not feel gritty,
erratic, or excessively stiff. If it does, loosen the slotted
lock ring, readjust the pivot-adjustment screw, and
retighten the slotted lock ring (as described above) until
the turret motion feels acceptable.
Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-5RE Modular Revolving Nosepiece
Figure
33).
Figure
Clean Off Any Visible Grease
Use a suitable solvent (e.g., mineral spirits) and a clean
rag or tissue to thoroughly clean any visible grease from
the exterior of the turret assembly. Be careful while
removing the excess grease, to prevent any of the
solvent from dripping into the revolving turret
mechanism and fouling the grease within.
Reinstall the Mechanical Detent Stop
Place the mechanical detent stop in position on the
stationary base, aligning the two holes in the
mechanical detent stop with the two tapped holes in
the stationary base (see
ball on the detent stop is facing downwards.
Figure 35 – Place the mechanical detent stop in position
Use a suitable JIS screwdriver to reinstall two M2X4
pan-head screws to secure the mechanical detent stop
in place on the stationary base (see
Figure 36 – Secure the detent stop with two screws
Figure
35). Make sure the stop
Figure
36).
Page 13 of 18
Revision 3