Olympus BX51M Руководство по эксплуатации - Страница 26

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Микроскоп Olympus BX51M. Olympus BX51M 36 страниц. System metallurgical microscope

Olympus BX51M Руководство по эксплуатации

SPECIFICATIONS

Item
Optical system
Reflected light illumination
Electrical system
Focusing system
Revolving nosepiece
Observation tube
Stage
Operating environment
23
UIS2 (UIS) (Universal Infinity System) optical system
Reflected light illuminator (BX-RLA2), tube magnification 1X, super widefield (FN 26.5) compat-
ible.
Available observations: @Reflected light brightfield
²Reflected light darkfield
³Reflected light Nomarski DIC
|Reflected light simplified polarized light
Rated input voltage: 100-120 V/220-240 V
Light intensity voltage range: 1.0 to 12.0 V DC (continuously variable)
12 V, 100 W long-life halogen bulb (pre-centered) 12V100WHAL-L (PHILIPS 7724)
12 V, 50 W long-life halogen bulb (pre-centered) 12V50WHAL-L (LIFE JC)
Average life time: Approximately 2000 hr. when used as directed.
Light preset switch: Preset voltage can be set arbitrarily between 1.0 and 12 V.
Power consumption: 140 W
Stage height movement by roller guide (rack & pinion)
Stroke per rotation: 0.1 mm (fine), 17.8 mm (coarse)
Full stroke range: 25 mm
Upper limit stopper
Tension adjustment on coarse focus adjustment knob.
U-5RE-2
DIC slider insertion type:
U-D6RE, U-D7RE, U-D5BDRE, U-D6BDRE, etc.
U-BI30-2
U-TR30-2
Widefield
Widefield
Widefield
binocular
trinocular
trinocular
Tube tilting angle: Fixed
Interpupillary distance adjustment: 50 to 76 mm
U-SIC4R2/SIC4L2
Coaxial knobs on bottom right (left).
4 x 4 inch stage
Drive system: Rack & pinion drive
Stroke: 100 Y-axis x 105 X-axis mm
Y-axis lock: Fixing in Y-axis direction
using a lock lever
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly
through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664)
· Installation (overvoltage) category: II (in accordance with IEC60664)
Specification
, 1.8 A/0.8 A, 50/60 Hz.
U-ETR3
U-ETBI
U-SWTR-3
Widefield
erected
erected tilting
widefield
binocular
trinocular
FN 22
Tilting angle
Tube tilting angle: Fixed
25°
Stage with coaxial knobs on the bottom right (left)
Drive system: Wire drive
Stroke 52 Y-axis x 76 X-axis mm
X/Y-axis knob rotation tension adjustment available
U-SWETR
U-SWETTR2
Super
Super
Super
widefield
widefield
erected
erected tilting
trinocular
trinocular
FN 26.5
Tilting angle
U-SVRM/USVLM
35°