Olympus CK30 Руководство по эксплуатации - Страница 30
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Микроскоп Olympus CK30. Olympus CK30 39 страниц. Culture microscope
SPECIFICATIONS
1. Optical system
2. Illumination
3. Focusing
4. Revolving nosepiece
5. Observation tube
6. Eyepiece
7. Objectives
25
Item
6 V 30 W HAL high-intensity halogen bulb
Philips 5761
(Average service life: Approx. 100 hrs. when used as directed)
Output rating: 6 V
Input rating: 100V area – – 100/110-120 V
Vertical movement of revolving nosepiece (stage is fixed)
Coarse/fine adjustment knobs (provided with tension adjustment mechanism)
Stroke (from focal point on stage surface): 7 mm upward and 2 mm downward
Binocular tube
Equivalent of CH3-BI45 incorporated in
CH3-BI45
microscope body
Trinocular tube
CH3-TR45
Tilting binocular tube
CK40-TBI
NCWHK10X
Field of view no. 18
WHK10X
Field of view no. 20
WHK15X
Field of view no. 12
For brightfield
EDAch4X
EDAch10X
LWDCDAch20X
For phase contrast
SPlan4XPL-6
DAch10XPL-6
LWDCDAch20XPL-6
LWDCDPlan40XPL-6
LWDCDPlan40XFPL-6
For relief phase
SPlan4XRP
contrast
DAch10XRP
LWDCDAch20XRP
LWDCDPlan40XRP
Specifications
CK30
LB optical system (finite-corrected system)
30 VA
200V area –– 220/230-240 V
Quadruple
N. A. 0.10 W.D. 29.0 mm
N. A. 0.25 W.D. 6.3 mm
N. A. 0.40 W.D. 5.4 mm
N. A. 0.13 W.D. 15.5 mm
N. A. 0.25 W.D. 7.18 mm
N. A. 0.40 W.D. 5.4 mm
N. A. 0.55 W.D. 2.04 mm with correction
collar
N. A. 0.55 W.D. 2.04 mm
N. A. 0.13 W.D. 15.5 mm
N. A. 0.25 W.D. 7.18 mm
N. A. 0.40 W.D. 5.4 mm
N. A. 0.55 W.D. 2.04 mm
CK40
0.6 A, 50/60 Hz
0.3 A, 50/60 Hz
Tube inclination: 45°
Interpupillary distance adjustment:
48 – 75 mm
Tube inclination: 45°
Interpupillary distance adjustment:
53 – 72 mm
Tube inclination: 30° – 60°
Interpupillary distance adjustment:
50 – 76 mm
NCWHK10X objective can only be used