SAFETY PRECAUTIONS
@
Fig. 1
7
Fig. 2
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols, and always use the equipment in
the safest possible manner.
Symbol
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury and/or
damage to the equipment.
l
Indicates that the main switch is ON.
\
Indicates that the main switch is OFF.
Warnings
Warning indications are placed where special precautions are required when handling and using the unit.
Warning indication position
1. Install the microscope on a stable, horizontal table. Make sure the table is
sturdy enough to support the microscope's weight. (Weight: Approx. 10 kg)
2. When transporting the microscope, always hold the observation tube
²
support @ and the stage handle ² using both hands. (Fig. 1)
# If the microscope is dropped under this condition, the focusing
mechanism may be damaged. To prevent this, insert the provided
6
protection pad ³ below the revolving nosepiece and fix the pad
using the coarse adjustment knob. (Fig. 2)
3. The surface of the lamp socket | on the illumination unit can get ex-
|
tremely hot. Make sure you leave a space of 10 cm or more around the
5
lamp socket, especially above and behind it, to dissipate heat. (Fig. 1)
4. When installing the microscope, route the power cord and lamp cord 5
away from the lamp socket | and illumination unit 6. Otherwise, the
power cord and lamp cord coverings may melt.
5. To avoid potential shock hazards and burns when replacing the bulb,
make sure that the main switch 7 is set to "
unplugged from the outlet, and that the lamp and the area around the
lamp socket | have cooled sufficiently. (Fig. 2)
Applicable bulb
3
6. Always use the power cord provided by Olympus. If no power cord is
provided, please select the power cord by referring to the section "PROPER
SELECTION OF THE POWER SUPPLY CORD" at the end of this instruc-
tion manual. If the proper power cord is not used, Olympus can no longer
warrant the electrical safety performance of the equipment.
7. Always ensure that the grounding terminal is properly grounded. If the
equipment is not grounded, Olympus can no longer warrant the electri-
cal safety performance of the equipment.
8. The standard service life of the lamp socket is 8 (eight) years of use or
20,000 hours of total power ON period, whichever is the shorter period.
For details, see Inspection Sheet on page 27.
Lamp socket
(U-LS30-4)
Halogen bulb:
6V30WHAL-L (Hosobuchi G4 20H CF-6) or
6V30WHAL (Phillips 5761)
Halogen bulb, 6V 30WHAL (Philips 5761)
Explanation
[Warning against high temperature]
GX41
" (OFF), the power cord is
1