Olympus SZ4045TR Руководство по эксплуатации - Страница 8
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Микроскоп Olympus SZ4045TR. Olympus SZ4045TR 16 страниц. Zoom stereo microscope
Fig.
1
Fig.
2
Diopter
adjustment
ring
Fig.
3
Left diopter
Right diopter
adjustment
ring
adjustment
ring
Fig.
4
(1) Usually,
the Stage plates are used with
the frosted
white
side facing
the objective.
However,
if the specimen
is white or brightly
colored,
use
the black side of the plate as the darkened
background
increases image
contrast.
(2) To remove the Stage plates, press down the supporting
end of the plate
with a fingertip,
and the other end will flip up. (Fig. 1)
(1) Turn one focusing
knob with
the other
focusing
knob tightened,
by
holding both knobs, in Order to increase or decrease the rotation tension,
depending
on the turning
direction.
(Fig. 2)
(2) This adjustment
is intended to prevent the microscope
body from drifting.
Set the knob tension
at a level slightly
stiffer
than that to insure no
unintentional
movement
and facilitate
easy Operation.
q
(1) Place the specimen in the Center of the Stage plate, and hold the specimen
with the Stage Clips if necessary.
(2) Illuminate
the specimen
using an optional
illuminator.
For SZ4045TR
(1) Minimize
the magnification
power by turning
the zoom control
knobs.
(2) Looking
through
the right eyepiece,
bring the specimen
into focus
by
turning
the focusing
knobs.
(3) Looking
through
the left eyepiece,
bring the specimen
into focus
by
turning
the diopter adjustment
ring. (Fig. 3)
(4) Maximize
the magnification
power
by turning
the zoom control
knobs,
and bring the specimen
into focus by turning
the focusing
knobs.
For SZ6045TR
and SZ1145TR
(1) Maximize
the magnification
power by turning
the zoom control
knobs.
(2) Looking
through
the right eyepiece,
bring the specimen
into focus
by
turning the focusing
knobs.
(3) Minimize
the magnification
power
by turning
the zoom control
knobs.
(Fig. 3)
(4) If the specimen goes out of focus, read just the focus by turning the right
diopter adjustment
ring only. (Fig. 4)
(5) Maximize
the magnification
power
again by turning
the zoom control
knobs.
If the specimen
goes out of focus,
repeat Steps (2) through
(4)
above. This allows more precise diopter adjustment.
(6) Minimize
the magnification
power
by turning
the zoom
control
knobs
and, looking through
the left eyepiece,
bring the specimen
into focus by
turning
the left diopter adjustment
ring only. (Fig. 4)