Adjusting the diopter (zoom confocality adjustment)

Confirm that the eyepiece clamping knobs a are
NOTE
tightened firmly before proceeding to the adjustment.
Eyepieces without eyepiece micrometer disk
1
Turn the diopter adjustment rings b of the left and right eyepieces to
positions "0". (This adjustment is not possible with eyepieces without
helicoids.)
2
Place an easy-to-observe specimen on the stage plate.
3
Set the zoom adjustment knob c for the lowest magnification and
rotate the focus adjustment knob d to bring the specimen into focus.
4
Set the zoom adjustment knob c for the highest magnification and
rotate the focus adjustment knob d to bring the specimen into focus.
5
Set the zoom adjustment knob c for the lowest magnification and
rotate the diopter adjustment rings d of the left and right eyepieces,
instead of the focus adjustment knob, to bring the specimen into focus.
Set the zoom adjustment knob c for the highest
TIP
magnification again and check the image focusing. The
diopter adjustment is complete if the image in accurate focus.
If not, repeat steps
Eyepieces with eyepiece micrometer disk
1
Look through the eyepiece with eyepiece micrometer disk and rotate its.
diopter adjustment ring b to bring the micrometer scale into focus.
2
Place an easy-to-observe specimen on the stage plate.
3
Set the zoom adjustment knob c for low magnification, look through
the eyepiece with eyepiece micrometer disk and rotate the focus
adjustment knob d to bring the specimen into focus.
4
Set the zoom adjustment knob c for the highest magnification,
look through the eyepiece with micrometer disk and rotate the focus
adjustment knob d to bring the specimen into focus.
5
Set the zoom adjustment knob c for the lowest magnification and
rotate the diopter adjustment ring of the eyepiece without eyepiece
micrometer disk, instead of the focus adjustment knob, to bring the
specimen into focus.
· Set the zoom adjustment knob c for the highest magnification
TIP
again and check the image focusing. The diopter adjustment
is complete if the image in accurate focus. If not, repeat steps
to
above.
3
5
· Note the left and right diopter scale values so that it can be
quickly duplicated in future observations.
to
above.
3
5
SZX7