17
Close the air outlet plug (pink) of the air jack by rotating the plug
clockwise.
18
Place the air jack on the caster on the front side. At this time,
confirm that the band of air jack is not removed.
19
Remove the cap of the air inlet plug (white) and supply air until stand F can be removed using the pump.
Stand F
CAUTION
Supply air in with the air jack in contact with the caster. Otherwise, the product may topple down.
NOTE
Remove stand F by supplying air while using the line (yellow) as a reference. If stand F is hard to remove, feed
more air with the pump.
Air jack
Figure 19
Air jack
Figure 20
– 7 –
Air outlet plug (pink)
Figure 17
Figure 18
Line (yellow)
Band
Air inlet plug
(white)
Pump