AccuPulse NIBP Simulator Руководства по эксплуатации - Страница 18

Просмотреть онлайн или скачать pdf Руководства по эксплуатации для Испытательное оборудование AccuPulse NIBP Simulator. AccuPulse NIBP Simulator 20 страниц. Hand held nibp simulator

Set Preset:
This selects a preset to simulate with.
Patient type 1=ADULT, presets are 1 through 7
Patient type 0=NEO, presets are 1 through 6
Send:
<SET_PRESET> <CR>
[0,1],[1-7] <CR>
Entered as "Patient Type" comma "Preset" with no spaces
Recv:
'=' – success
'#' – preset out of range.
'$' – checksum error in data that was sent. Preset is not valid in the current
CalTable.
Set UTP
User Defined Test Points is an optional software module.
Sets the index of the User Test Point. Valid range is 1..45
Send:
<SET_UTP> <CR>
[1..45] <CR>
Recv:
'=' – success
'#' – out of range.
Set PR
Set Pulse Rate from 15 to 330 BPM.
Send:
<SET_PR> <CR>
[15..330] <CR>
Recv:
'=' – success
'#' – pulse rate is out of range.
Set AMP
Configure the Pulse amplitude from 0 to 150% of normal.
Send:
<SET_AMP> <CR>
[0..150]
<CR>
Recv:
'=' – success
'#' – amplitude is out-of-range.
Set SHIFT
Sets the amount to shift the entire envelope, in mmHg
Send:
<SET_SHFT >
[-50..100] <CR>
Recv:
'=' – success
'#' – shift value is out-of-range.
'$' - shift was applied but at a reduced value to remain within simulation
limits.
Set CalTable
CalTables are an optional software module
This command selects an optional CalTable for use.
<CR>
The default 'Generic' is number 46 and is provided in all NIBP units. Numbers 1-45 can
be populated with CalTables created by Clinical Dynamics to match specific equipment.
Currently the first 10 are used.
Send:
<SET_CALTABLE> <CR>
[1..46] <CR>
Recv:
'=' – success
'#' – out of range.
Get CalTable
Send:
<GET_CALTABLE> <CR>
Recv:
xx
'='
Note: CalTable index is valid only in BP Mode
Get BP Parameters
Send: <GET_BP_PARAMS> <CR>
Recv:
PRESET_TYPE, BP
PATIENT_TYPE, 0
PRESET, 3
SYS, 80
DIA, 50
PR, 80
SHIFT, 15
AMP , 100
DSYS, 0
DDIA, 0
Configure Leak Test:
While in Leak Test Mode, this sets the start pressure and time for Leak Test
Pressure range is 50 through 300 mmHg
Time range is 20 through 120 seconds
Send: <LEAK_CFG> <CR>
[50...300],[20...120] <CR>
Entered as "Pressure" comma "Time" with no spaces
Recv: '=' – success
'#' – out of range.
To begin the test Send:
<BUTTON>
F3
Auto Leak Test: Setting Test Pressure...
=
>AccuPulse LeakTest Results:
Start Pressure= 299 mmHg
119 Sec. Pressure Drop= 2 mmHg
End of test results are reported as shown above.
18
[ BP or UTP ]
[ 0=NEO , 1=ADULT ]
[ preset within patient type, 1-7 Adult, 1-6 Neo, or 1-
45 optional UTP mode ]
[ base setting without any shift value ]
[ base setting without any shift value ]
[ BP Presets only, not returned when using optional
UTP mode ]
[ Generic Trim value, not returned when using
another CalTable ]
[ Generic Trim value, not returned when using
another CalTable ]