FHC STar 66-MP-EH Руководство по применению - Страница 9

Просмотреть онлайн или скачать pdf Руководство по применению для Адаптер FHC STar 66-MP-EH. FHC STar 66-MP-EH 9 страниц. For microtargeting platforms

FHC STar 66-MP-EH Руководство по применению

Track Selection

Center positioner (9 distinct tracks): Center track on target. Four parallel tracks with 2mm offset,
'+' configuration, using positions A, C, E and G. Four parallel tracks with 2mm offset, 'x' configuration,
using positions B, D, F and H.
Examples:
Center Positioner, '+'
configuration, using any
of positions A, C, E or G,
indicated by the black
dots in the offsetting chart
below.
3mm offset positioner (40 distinct tracks): Five parallel tracks with 2mm offset, center track offset
3mm from origin, in eight configurations using positions A-H.
Examples:
3mm Offset Positioner in
the 'G' position, indicated
by the dark blue dots in the
offsetting chart.
Offsets beyond the 10mm range of the positioner may be reached with the target offset hub.
Track Offsetting
Chart
3.5
3.0
2.12
2.0
1.41
1.0
.71
.71
1.0
1.41
2.0
2.12
3.0
3.5
L011-44
Center Positioner, 'X'
configuration, using any
of positions B, D, F or H,
indicated by the red dots in
the offsetting chart.
3mm Offset Positioner
in the 'B'
position, indicated by
the magenta dots in the
offsetting chart.
Rev. E1
2019-05-13
WARNING: The target offset
hub creates a new trajectory
with target area orthogonally
offset 6mm from the original at
the microelectrode tip depth
when the physical drive reads
30mm. For platforms with
T<>30, the offset of 0.69mm
per 10mm of drive travel may
be used to determine position
at target depth.
WARNING: The entry
offset hub creates a new
trajectory which is orthogonally
offset 3mm from original at
approximate entry depth, and
coincides with the original at
the microelectrode tip depth
when the physical drive reads
30mm. For platforms with
T<>30, the offset of 0.34mm
per10mm of drive travel may be
used to determine position at
target depth.
Page 9 of 9