ABB 10D1476 Руководство по установке - Страница 5
Просмотреть онлайн или скачать pdf Руководство по установке для Измерительные приборы ABB 10D1476. ABB 10D1476 10 страниц. Magnetic flowmeter
MOUNTING PROCEDURE
INSTALLATION OF 1/10 - 3/8 INCH METERS
• Align flange adapter on meter
body, as shown below.
• Install a gasket at meter inlet and
outlet, as applicable for meter
type. Observe flow direction arrow.
• Insert meter with (with gaskets)
between pipe flanges.
• Insert mounting studs through bolt
holes.
• Thread nuts on mounting studs
(finger tight).
• Press down on flange adapter to
force-spread studs, thereby cen-
tering meter in pipeline.
• Use proper torque and tightening
sequence (i.e. 1-3-2-4) to maintain
even
pressure
shown in the figure to the right.
ABB
FIGURE 4. Meter Mounting Diagram, 1/10 - 3/8 in. Sizes
INSTALLATION INSTRUCTIONS
distribution
as
3 - 10
(1/10 - 3/8)
NOTE: Meter shown is Model 10D1476/1477
4