PACHYMETER

How To Measure A New Patient

1.
Step One: Begin a new patient, by selecting
Patient.
3.
Step Three: Launch B-Scan Mode by clicking the
Pachymeter
icon.
5.
Step Five: Begin measurement by clicking the
Reading
icon.
8 | To order in the U.S. and Canada call (800) 979.2020 • (610) 889.0200 International • +31 (0) 485.350300 Europe or visit
New
2.
Step Two: Type the patient data into the appropriate
fields.
4.
Step Four: On the measurement screen, choose the
eye (OD/OS) and then choose the scan location on
the cornea.
Start
6.
Step Six: Place probe on the eye in the location
selected on the screen. The flat portion of the probe
tip should be completely in contact perpendicular
with the patient's cornea (not at an angle).
Measurements will not be obtained unless the probe
is in complete contact with the cornea.