AccuTools eL-320 Руководство пользователя - Страница 5

Просмотреть онлайн или скачать pdf Руководство пользователя для Датчики безопасности AccuTools eL-320. AccuTools eL-320 6 страниц. Cli020202 combustible gas leak detector

RECALIBRATION
When the eL-320 is turned on, the unit automatically calibrates
the sensor for maximum sensitivity. Occasionally, the signal
from the sensor may drift (especially if moving from one location
to another where the ambient humidity changes), causing a
continuous alarm that will not clear. In this case, simply power
the unit off and on again, or use the following recalibration
procedure:
To recalibrate the unit:
1. Move the unit away from any suspected leaks and into fresh
air.
2. Quickly press the button four (4) times. This must be done
within one second.
3. The LED will flash rapidly for about 10 seconds, indicating
recalibration.
4. The unit will return to normal operation in the high sensitivity
mode.

BATTERY INSTALLATION AND REPLACEMENT

1. Turn the power off.
2. Remove the battery cover from the back of the instrument by
loosening the battery cover screw and lifting the battery
cover off.
3. Insert 2 size 'AA' alkaline batteries, observing proper battery
polarity.
4. Replace battery cover and tighten battery cover screw.
eL-320 User's Guide
5