IV. DISS HOSES
1. Attach one end of the blue hose with DISS connector to the DISS fitting at the far left side of the
bottom of the main manifold. Attach the other end of the hose to the fitting on the nitrous oxide
regulator.
Note: If system is a 2 x 2 manifold, in the same manner attach the second blue hose to
the second open fitting at the far left side of the bottom of the main manifold. Attach
the other end of the hose to the open fitting on the nitrous oxide regulator.
2. Attach one end of the green hose with DISS connector to the DISS fitting closest to the center of
the bottom of the manifold. Attach the other end of the hose to the fitting on the oxygen regulator.
3. Attach the second green hose to the open fitting on the manifold and attach the other end of the
hose to the fitting on the oxygen regulator.
Note: Use 2 wrenches when tightening the fittings.
V. DESK MOUNT ALARM
Desk Mount Alarm Installation - Models #38003, #38005 and #34065
1. Take the desk alarm unit, the cable connection receptacle box and the receptacle box faceplate from
the shipping carton and remove their plastic wraps.
2. Select an appropriate location for the desk alarm unit.
3. Attach receptacle box to wall studs.
Note: The receptacle box and the desk alarm unit must be located with 12 feet
of each other.
4. Thread the gray cable through one of the knockout holes in the receptacle box. Keep 6 to 8 inches
of the end of the cable coiled inside the box.
5. To prevent joint compound and other debris from getting into the receptacle box during building
construction, cover the box opening with masking or duct tape (seal tape at receptacle box mud
ring).
6. Run the length of the gray cable to the tank room. Coil the end of the hose a minimum of 6 to 8
inches at the end of the oxygen gas line in the tank room.
7. Anchor the running gray cable to the studs and to the frame of the building.
CAUTION: DO NOT NAIL OR STAPLE THROUGH THE CABLE. ANY BREAK IN THE CABLE'S
INSULATION MIGHT COMPROMISE THE SYSTEM'S INTEGRITY AND CAUSE THE ALARM
TO NOT OPERATE PROPERLY.
Note: The gray cable is 100 ft long. Contact Accutron, Inc., at 800-531-2221 if additional
cable is required.