Checking the programmed values
NOTE
Potentiometer
Fig. 3
Prerequisites
T
The
w
Connect the service PC to the serial interface of the SIREMOBIL .
Start the service program.
Select the Adjustment, Calibrations menu.
Click "Yes" to answer the query.
In the Combo box "Function Group:", select "Diamentor".
In the window "Available Functions" , select the line "Start Diamentor Test" and click on
"Execute" .
Close the pop-up window with the display "Result 0" by clicking the OK button.
The tested
This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: +/- 5 %.
By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
The potentiometer (fig.2) of the area dose product measurement device can be used to
adjust the test value to the programmed value (Qt*Kg).
SIREMOBIL Iso-C
Area dose product measurement device
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
product measurement device.
value is already programmed. Refer to Programming.
T
value is displayed on the monitor of the SIREMOBIL .
w
SPR2-230.033.04
Rev. 03
06.05
Page 12 of 16