PROXIMITY FUNCTIONS:
Pressing the "Proximity" (Lights BLUE LED) is activated on / off the motor by an infrared proximity sensor that detects the
presence of the operator's hand from a few inches away. This allows the operator to use the device without touching or focus
on pressing the button.
The on / off button is also active with the function of proximity inserted and can be used as an alternative.
To remove the "Proximity" you have to re-press the button.
The "Proximity" is retained, or if it was active before power off, then on again to return to this, but if it was not active will remain
disabled. The function set to power down the card after 20 minutes after turning off the engine if it is not then on again.
The unwanted approach of the hand to the Proximity sensors causes the device to switch off. To reactivate the
function, place the hand close to the sensors.
NOTE: It is up to the end user to activate or deactivate the Proximity sensor. With the Function off, the device is activated/deactivated
by pressing the ON/OFF key.
LIGHT INDICATORS:
The device is provided with a light indicator (directly on the front panel) which allows you to view the operation of the device,
autonomy of the battery, and the charge phase in progress.
The light indications, which appear during operation, are indicated in table I.
TAB. I – INDICATOR LIGHTS DURING OPERATIONS
LED signals
Flashing Green Led
Steady Greed Led
Steady Red Led
Flashing Red Led
Steady Yellow Led
TAB. II – WRITTEN SIGNALS / BUTTON LED SIGNAL
Button LED signal /
written signal
ON/OFF button
Proximity Button
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the blu ring fitted on
the tank holder for a more comfortable insertion of the same container.
After opening the package, fully stretch the bag and then flatten it concentrically to eliminate as much air as possible.
-
Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing firmly around the entire
-
perimeter. Make sure that the system is completely sealed.
close the connector marked as "TANDEM" with the lid provided.
-
Connect the power source of the vacuum to the VACUUM port equipped with specific reusable conical fitting with "male"
-
connection.
Connect the patient tube to the PATIENT port of the cover
-
Before use, check all closures and make sure there are no leaks, starting the aspiration source. If the bag expands to fully
-
adhere to the walls of the rigid container and the cover bends towards the inside of the glass, the system is not leaking.
Start the aspiration and periodically check the filling level of the container. The overflow valve will cause the interruption of
-
aspiration if the aspirated fluids have reached the maximum filling level of the device.
When the float valve intervenes signalling the device is too full, the suction source must be disconnected within no more than
-
5 minutes.
Warning: The accidental inversion of connections may cause contamination for the operator and/or for the vacuum
generation equipment.
Phase
During rechanrge
During recharge
During battery
operation
Device automatically
turns off when the
battery is flat
During battery
operation
Function
Power on
Switch the ON/OFF key
Problem / Cause
Battery recharge running
Recharging cycle complete
Flat battery
Battery completely flat
Intermediate status
Colour
Green / Yellow / Red
Blue
9
Solution
Wait
Remove the device from the
support plate (ASKIR SUPPORT
118 - mod. ASU 118) of the
ambulance
Start recharging cycle
WARNING: During this signal, you
will hear a long, continuous beep
(duration of sound 0.8 sec / sound
frequency: every 8.5 sec, which
notifies the user regarding the
battery discharge.
When the device is restarted the
LED will flash red: begin the
battery recharge cycle
immediately
Guaranteed battery function /
Recharge when the red LED signal
comes on.
Position
Near the front panel key
LED located above the
"Proximity" key