3000 A RANGE
Primary Current 150 A
Accuracy %
Phase shift
Load Impedance: ≤0.4 Ω
Overload: 3600 A for 10 min
Ampere-Second Product: 80 A·s
Accuracy: Per IEC 185-26-27
Class 0.5 (5 V·A) from 48 Hz to 1000 Hz
(*Reference conditions: 23 °C ± 3 K, 20 %RH to
85 %RH, 45 Hz to 65 Hz, external magnetic field
<40 A/m, no DC component, no external current
carrying conductor, test sample centered.)
Frequency Range: 30 Hz to 1500 Hz; cur-
rent derating above 1500 Hz for continuous
use
Load Impedance: ≤2 Ω
Working Voltage: 600 V
Common Mode Voltage: 600 V
Influence of Adjacent Conductor:
<0.005 A/A
AC
Influence of Conductor in Jaw Opening:
1.5 % ± 0.2 A on 1000:5 range
1 % ± 0.2 A on 2000:5 range
1 % ± 0.2 A on 3000:5 range
MECHANICAL SPECIFICATIONS
Operating Temperature:
14 °F to 122 °F (-10 °C to 50 °C)
Storage Temperature:
-13 °F to 176 °F (-25 °C to 80 °C)
Influence of Temperature:
<0.1 % per 10 K
Jaw Opening: 3.54 in (90 mm)
Maximum Conductor Size:
Cable: 2.52 in Ø max (64 mm)
Bus bar: 1.97 in x 5.31 in (50 mm x 135 mm)
2.52 in x 3.94 in (64 mm x 100 mm)
OPERATION
Please ensure that you have read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model JM865A
Connect the black and red terminals to the Ampere AC range of your DMM or current
measuring instrument. Select the appropriate current range. If the current magnitude
is unknown, select the highest range (3000 A
the handle. Clamp the probe around the conductor with the arrow pointed toward the
load. Select the lower range until you obtain the best resolution. Read the value display
on the DMM and multiply it by the range selected. If the reading equals 2.59 A on the
1000/5 range, the current flowing through the probe is 518 A
For the best accuracy, carefully center the conductor inside the probe jaw, avoid if
possible, the proximity of other conductors which may create noise.
600 A
3000 A
1.5 % 0.75 %
1.5 °
0.75 °
AC
AC
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
0.5 %
1 m (IEC 68-2-27)
0.5 °
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10 Hz, 0.15 mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10 % fiberglass charged
Jaws:
Dimensions:
4.72 in x 12.40 in x 1.89 in
(120 mm x 315 mm x 48 mm)
Weight:
2.65 lb (1200 g)
Colors:
Dark gray handles with red jaws
Output:
Two standard safety banana jacks (4 mm)
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between primary or secondary and outer
case of handle upon IEC 1010-2-32
Common Mode Voltage:
600 V CAT III, Pollution: 2
ORDERING INFORMATION
Current Probe JM865A ........ Cat. #2110.88
Accessories:
Adapter - Banana (Female) to BNC (Male)
(XM-BB) (600 V CAT III) ........ Cat. #2118.46
Lead - set of 2, 5 ft (red/black) PVC
with 4 mm ST/ST plugs ......... Cat. #5100.07
/ 5 A
AC
polycarbonate UL 94 V0
ABS UL 94 V2
) on the switch probe inside
AC
(2.59 A x 200).