Crowcon C-Test Руководство пользователя - Страница 10

Просмотреть онлайн или скачать pdf Руководство пользователя для Испытательное оборудование Crowcon C-Test. Crowcon C-Test 12 страниц. Gas (bump) test and calibration station for use with the crowcon clip and crowcon clip+

Crowcon C-Test User Manual
7.9.

Monitor Settings

Users can program Crowcon Clip & Clip+ configurations for
each
sensor
type.
programmed to the C-Test, every monitor that is bump tested
will take on the new configuration values.
In order to make changes to any particular field please update
the value and ensure the check box on the left hand side is
checked. Below is a description of the individual fields:
User ID: Each individual unit can now have a unique electronic
user ID. You are able to program in up to 6 characters. Note:
Some alphabetic characters do not display well.
Calibration Interval: In order to help with Calibration test
accountability users can adjust the Calibration Interval to
company requirements. If a unit has not been Calibrated within
the time frame the unit will prompt the user to run a Calibration via the C-Test.
Bump Interval: In order to help with bump test accountability users can adjust the Bump Interval. If a
unit has not been bump tested within the time frame the unit will emit a flash until it has been tested
either manually or via the C-Test.
Self Test Interval: By default all Crowcon Clip & Clip+ instruments have a default 20 hour self test.
Frequency can be set anywhere from 8 to 20 hours.
High Alarm/Low Alarm Set Points: Adjust the high/low alarm set point. Note: When doing so make
sure to set the alarm limit not to exceed the local or site guidelines. Some sites may have different set
points than others.
Show Sensor Readings: Once the unit starts to see gas the unit will display the real time reading with
a 0.1 ppm resolution. If this is unchecked the unit will always display the life remaining or flash low/high
alarm in an alarm condition
Hide bump due LED Flashing: If you prefer to not have the units flash when due for a bump test,
please make sure to check this box.
Hide Clock: This will turn the monitor into a real time ppm display monitor. By default always showing a
real time ppm reading.
After you have made your requested changes click on the Write Device Button. If a successful write has
occurred you will see a green bar at the bottom (see above).
Firmware updates: The C-Test allows for firmware upgrades to occur during the lifetime of the unit.
These may be product improvements to add additional features or to enhance monitor performance.
Most firmware upgrades are not required unless notified by Crowcon. Once new firmware has been
programmed to the C-Test, units will be upgraded before testing. During the upgrade the LED lights will
remain orange in colour and flash. DO NOT move the unit during a firmware upgrade.