4.7 RS232 Menu [.rs232.]
Baud [BAUD] – The RS232 baud rate can be set to 600, 1200, 2400, 4800, 9600 and 19200.
Parity [Parity] – Parity can be sent to 7 bits-even parity [7 eveN], 7bits-odd parity [7 Odd],
7bits-no parity [7 NO] or 8bit-no parity [8 NO].
Handshake [Handsh] – Handshake can be set to off [Off], X on – X off [on-off], or hardware [HardwWr].
4.8 Lock Menu [.LOC.]
When a Lock Menu item is SET ON the indicated menu cannot be changed.
[Loc Cal] – Calibration, [Loc SEt] – Setup, [Loc.mMod] – Mode, [Loc.Unit] – Unit,
[Loc Prt] – Print, [Loc 232] – RS232.
4.9 Sealing Access to the Balance Settings
The Menu Lock switch prevents changes to the Lock Menu. The switch can be secured using paper seals,
wire seals or plastic ties.
Un-Locked
Locked with Plastic Tie
4.10 Approved "M" Models
Balances marked with an "M" have been calibrated and sealed at the factory for use as an EEC approved
balance. The Cal, Setup, Mode and Unit menus will be locked.
4.11 Local Approval
The InCal models can be approved by local weights and measures authorities. The calibration must be
verified and the Menu settings must be set appropriately before being locked using the Lock Menu. The
Menu Lock switch on the bottom of the balance can then be sealed in the locked position. The supplied
capacity label is to be placed below the display.