3.5.4

Passive Testing

Passive testing is a test that is performed in a non-invasive manner and without using
mouthpieces. Passive testing on the LE5 can activated in one of two ways, either button
activated by the operator or pressure activated by the donor blowing.
A passive test by button activation is performed by holding the instrument 5 cm in front of
the test subject's mouth, Press the Test Key and ask them to count or talk so that exhaled
breath may be picked up by the instrument and analysed for the presence of any alcohol.
Pressure Activated Passive is performed by holding the instrument approximately 5cm in
front of the test subject mouth and asking them to blow into the Passive Cup. The pressure
of the blow activates a passive test.
When using Passive Standard testing, a Mouthpiece test is automatically enabled, should
the instrument detect alcohol on the subject's exhaled breath. Passive Only testing does not
prompt the user for further testing.
3.5.5

Enabling Passive Standard

If Passive standard mode has not been enabled, proceed as follows:
• With the screen displaying press to test, press the Up and Down buttons
simultaneously. The User Menu will be displayed.
• Press the down button to scroll down to Test Type.
• Press the Test key until Passive Std is displayed (Refer Figure 13).
• Passive Standard is now enabled. To leave the user menu, press the Up and Down
buttons simultaneously or press the Down button until Back is highlighted followed
by the Test key.
3.5.6

Taking a Passive Standard Breath Sample

If any step in a breath test sequence is not completed or the instrument is left idle for
longer than 60 seconds, the instrument will terminate the breath test and display test
timed out. The instrument will return to the press to test display.
If a test time out does occur during a breath test, a new breath test must be performed.
Perform a Passive Standard test as follows:
REF: Alcolizer LE5 Alcohol Tester User Manual V15 docx
Figure 13- Passive Standard Displayed
Note