1 Prepare EC Calibration Station
• Turn on gas at cylinder
• The gas value is determined using value on side of Ethanol bottle.
• The gas bottle setup is done by a senior technician.
2 Prepare the computer program
Turn on computer and connect to EC Calibration Station found on the desktop.
3 Remove EC from packaging from customer and restock for dispatch
• Check customer details on returned Calibration Request form (if form not returned
by customer complete a blank form with details to keep with paperwork).
• Get replacement Calibration Request Form (these forms are currently printed by
Print Today if stocks are running low please advise Administration Manager).
• Get replacement red/white Australia Post satchel with return AT address label.
• Fold Cal Request Form & Satchel together and neatly place into the EC Manual on
bottom of case.
• If unit is returned by customer in its hard case replace any missing mouthpieces
If the unit is an NRMA unit a disclaimer note should be included in the box. (Disclaimer is located at
S:\Charity\4 Forms\NRMA Disclaimer.doc). This applies to new units only.
4 Prepare EC for Calibration
• Remove EC from soft pouch
• Visually inspect unit for any damage record and damage in the notes section.
• Remove calibration sticker from back of unit. Attach to recal request form – in the
comments section.
• Unscrew battery cover.
• Unscrew main holding screw located under battery cover
• Record Fuel Cell date – on bottom of recalibration form
• Check black foam is in place on inside front cover – if not placed in correct position
(refer to assembly work instructions if necessary)
• Remove fuel cell from connection pins and check for corrosion. If no corrosion,
reposition fuel cell. If corrosion is evident, clean pins and replace fuel cell if required.
• Check that the pump volume is set between 0.11 and 0.15, 0.13 typical. Adjust and
set if required.
• Inspect pump tubing for any contamination and replace if required.
• Screw cover back together. Clean case if required.
• Replace 2 x AAA batteries.
• Perform a Current Drain test by removing one of the batteries. The unit should stay
on. If the screen fades, there is a current drain. This will require a new board.
REF: EC Service and Calibration Instructions V8