Alemite 3641 Руководство по эксплуатации - Страница 8
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Блок управления Alemite 3641. Alemite 3641 14 страниц. Electronic preset metered control valve
SER 3640
Recalibration
It may become necessary to recalibrate the meter.
Variables that cause a meter to require recalibration are:
• fluid viscosity
• fluid temperature
• flow rate
NOTE: The original scale factor is recorded
by the factory on the back of the lever. See
Figure 2-B.
Scale Factor Calculation
IMPORTANT: Dispense the fluid at the nor-
mal flow rate in the Manual mode of opera-
tion.
1. Dispense any amount of fluid greater than 1 gallon
(3.8 liters) into an appropriate-sized graduated beaker.
2. Record the amount dispensed and the value displayed
on the meter.
3. Insert a blunt tool (5/32 "allen wrench) into the access
hole in the back of the meter.
• See Figure 4.
4. Press and hold the tool for 2 seconds.
• The screen begins to flash and then displays the
current scale factor and unit of measure.
• The current unit of measure flashes.
5. Record the current scale factor.
6. Divide the value of the amount dispensed by the value
displayed on the meter. Multiply this product by the
existing scale factor. The product is the new scale
factor.
EXAMPLE: Exactly 4 quarts were dis-
pensed into the beaker and the meter regis-
tered 4.16 quarts. If the existing scale factor
is 0.957, a new scale factor of 0.920 must be
entered.
(4 ÷ 4.16) x 0.957 = 0.920
Revision (10-04)
Electronic Preset Metered Control Valve
Change the Scale Factor
7. Press the RESET button.
• The first digit of the scale factor flashes.
8. Press the TOTAL button until the required value
appears.
NOTE: The first digit can be set to 0 or 1.
All remaining digits to the right of the deci-
mal point (comma) have a full range from
0 to 9.
9. Press the RESET button.
• The next digit of the scale factor flashes.
10. Repeat steps 8 and 9 for all digits.
11. Once the last digit is set, press the RESET button.
• The pulse delay factor (PS) screen appears.
• See Figure 6.
IMPORTANT: The pulse delay factor should
always be set to zero. No pulse delay is
required on these control valve models.
12. Insert the tool into the access hole in the back of the
meter once again.
13. Press and hold the tool.
• The screen flashes three (3) times and then goes
blank.
14. Press the RESET button.
• The normal delivery screen appears.
Figure 6 Pulse Delay Factor Screen
8
Alemite Corporation