TEST PROGRAM
© TEST-OE : Checking the measuring optical system adjustment
Press this
key to quit
TEST-OE.
• TEST - OE can be used to test the accuracy of adjustments made on the measuring
optical system (Image rotator and photo sensor)
• Model eye alignment
Place the model eye tool on the centering tool and align it.
Focusing : Adjust the focus by pushing or pulling the J35058 model eye
attaching tool which installs the model eye.
Alignment: While observing the alignment mark, make a rough alignment of the
model eye position by turning the centering knob of the cenrering tool
until the asterisk " * " mark is positioned near the center of the
display.
Then make a fine alignment while observing the "XAL: " and
"YAL:" displays to obtain "XAL: +0.00" and "YAL: +0.00"
respectively.
Make sure that "TOT: 093" or a similar value is also displayed. If
this value is not displayed, adjust the VR3 GAIN control.
• Measurement
When you press the measurement button, the Image rotator rotates and
measurement starts. The asterisk
mark in the alignment mark moves around
and the "XAL: " and "YAL: " values change as the Image rotator rotates.
• Displaying the measurement results
The NR - 5500 displays the total of the alignment offsets measured at 72 measuring
points at the 5 degrees steps.
[X][Y]: Totals of alignment offsets in X and Y directions
At checking
: 50 or less
At adjustment: Adjust so that the total will be 30 or less.
X
: Total of alignment offsets in X direction only
Y
: Total of alignment offsets in Y direction only
* Calculation formula
[X]
[YJ = -rr-r
X
( | X L
J
+ | Y i | )
0 4
i»l
X
= 77
Z
Xi
(Signed)
6 4
i-i
Y
= -TT-r
X
Y i
(Signed)
6 4
i-i
(6-1 n