Abbott i-STAT 1 Краткое справочное руководство - Страница 5
Просмотреть онлайн или скачать pdf Краткое справочное руководство для Измерительные приборы Abbott i-STAT 1. Abbott i-STAT 1 7 страниц. For use with a clia certificate of waiver
Также для Abbott i-STAT 1: Руководство пользователя (9 страниц)
Patient Testing Procedure
Prepare the Sample
• Mix the sample tube by gently inverting it 2-3 times before filling a
cartridge. Do not use a partially filled collection tube.
• CHEM8+: Test the sample within 10 minutes after collecting it.
• 6+, EC4+, E3+, Crea, and G: Test the sample within 30 minutes
after collecting it.
Test Sample
1. Press
2. Press
3. As prompted by the handheld, use number
keys to enter the Operator ID and Patient ID.
4. Scan the barcoded cartridge lot number.
5. Remove cartridge from pouch.
6. Use a sample transfer device to remove some
sample from the tube. Dispense 3 drops of
sample onto a gauze pad.
7. Hold the sample transfer device over the
cartridge sample well.
8. Dispense sample until blood reaches the
cartridge fill mark. Some blood must be left in
the cartridge well.
Fill cartridge
to turn on handheld.
2
(i-STAT Cartridge).
Get sample
Continued
5