The meter provides a number that increases with the amount
of magnetite present. The following aspects should be noted:
1. The meter is corrected for gravity but it is good practice to
zero the meter, away from the pipe, in the same
orientation as the measurement to follow.
2. The meter is designed to work with a space between the
probe cap and magnetite. If the meter is too close to
magnetite (or indeed a ferritic sample) then a raw reading
of 2047 will be displayed.
3. Warning, magnetite distribution will effect the measured
value i.e. a small amount of material close to the probe
will give the same signal as a larger amount further away.
Magnetite tends to accumulate on the corners of cooling pipes
where the pipes turn from vertical to horizontal. At these
points the pipe fills up from the bottom rather than
homogeneously as per our occlusion calculations.
Under these conditions we would recommend probing the
underside, the side and the top of the pipe around the
circumference. Due to the proximity of other pipes the
underside and top of the pipes may not be accessible so we
recommend probing positions A, B and C in the Figure 2.
As the pipe fills up the signal is expected to increase on A first
then B and finally C.