To Adjust Settings
Press Menu
Press Desired
Setting Button
Setting
Default Setttings
For Forward and
Recommended setting
Reverse
for oscillation/tissue
resection
Troubleshooting: Faults and Solutions
Fault
Unit does not power on
Handpiece does not work
Motor runs but shaver
blades/burrs do not work
No suction
System runs with the
control unit but not with
the footswitch
Connected handpiece
not shown on display
This description of technique is provided as an educational tool and clinical aid to assist properly licensed medical professionals
in the usage of specific Arthrex® products. As part of this professional usage, the medical professional must use their professional
judgment in making any final determinations in product usage and technique. In doing so, the medical professional should rely on
their own training and experience, and should conduct a thorough review of pertinent medical literature and the product's Directions
For Use. Postoperative management is patient specific and dependent on the treating professional's assessment. Individual results
will vary and not all patients will experience the same postoperative activity level and/or outcomes.
www.arthrex.com
© 2019 Arthrex, Inc. All rights reserved. LI1-000000-en-US_A
For Footswitch
Override:
Select desired option
and then do one of the
following:
Press HOME to save
setting and return to main
screen.
-OR-
Press Setting Button to
save setting and return to
MENU.
-OR-
Press X to return to
MENU without
changing default.
Cause
1. Not receiving power from the wall receptacle or the
power strip.
2. Blown fuses.
3. Console has an internal power failure.
1. Handpiece is not connected to the correct channel.
2. Not fully or properly connected or connector is damaged.
3. Liquid in the receptacle or connector.
4. Connector is damaged.
5. Footswitch is defective.
6. Is not a cause for malfunction.
7. Handpiece is too hot after sterilization.
8. Handpiece over-current is detected.
9. Defective handpiece.
1. Shaver blade/burr incorrectly inserted.
2. Damaged hub.
3. Bent sleeve.
1. Clogged suction connection.
2. Clogged shaver blade or burr.
3. Closed suction control.
4. Damaged suction control.
1. Not fully or properly connected or connector is damaged.
2. Liquid in receptacle or connector.
3. Defective footswitch.
1. Not fully or properly connected or connector is damaged.
2. Defective console port.
3. Defective handpiece.
For Oscillation
For Language:
Mode:
Select desired language
Select the associated
and then do one of the
handpiece that is plugged
following:
in. Choose from Standard,
Press HOME to save
Efficient, or Aggressive
setting and return to main
mode. To save your
screen.
selection, press the home
-OR-
button.
Press Setting Button to
-OR-
save setting and return to
Press X to return to
MENU.
MENU without
-OR-
changing a selection.
Press X to return to MENU
without changing default.
Solution
1. Check the main power plug and wall receptacle.
2. Check the fuses.
3. Send in for repairs.
1. Check the connector is plugged into the correct channel
receptacle.
2. Check both connectors are plugged in properly.
3. Dry thoroughly.
4. Send in for repair. The account cannot replace the cable.
5. Operate the device using the console unit.
6. Replace the handpiece.
7. Allow it to cool down as indicated in the WARNING in
Section 7.0.
8. Allow 30 seconds for the console to reset the affected
channel.
9. Send in for repairs.
1. Insert the shaver blade so that it engages. Fully close
the locking mechanism.
2. Replace the shaver blade/burr.
1. Rinse with a syringe.
2. Clean with a brush.
3. Open suction control.
4. Send in for repairs.
1. Check the connector from the footswitch to the control unit.
2. Dry thoroughly.
3. Send in for repairs.
1. Check that both connectors are plugged in properly.
2. Check the connector with another handpiece.
3. Send in for repairs.