Extract from the
Operating Instructions
Hettich ROTOFIX 32A
Centrifuge Preventative Maintenance Schedule
Taken from the operator manuals supplied with each centrifuge
!
Pull the mains plug before cleaning
Using detergents follow the instructions of the Hettich Operating Instructions.
Centrifuge Bowl Seal
Cleaning
regularly
Rotor
Cleaning
at least once a week
Buckets
Cleaning
at least once a week
Motor Shaft
Cleaning
at least once a month
Surfaces and Chamber
Cleaning
weekly
Most Common Error Codes and Clearing Procedures
ERROR CODE – 1 (tachometer error)
Clearing Procedure:
• Wait for rotor stand still (120 seconds after rotor stops)
• Turn the power off and wait 10 seconds
• Turn the power on
• Check to make sure rotor is installed properly
before spinning
ERROR CODE – 2 (mains interrupt)
Clearing Procedure:
• Wait for rotor stand still
• Open the lid
• Press the start button
• Close lid and start again
ERROR CODE – 3 (balance error)
Clearing Procedure:
• Wait for rotor stand still
• Check to make sure the centrifuge is sitting on a level
and sturdy surface
• Make sure the rotor is installed properly and the bucket
weight is distributed evenly before spinning
Rubber Maintenance
after cleaning
Lube
Secured
after cleaning
weekly
Lube
after cleaning
Lube
at least once a month
Maintains good lab practices for cleanliness. Always inspect for damage.
Maintains long life for seal.
Always inspect seal for damage.
Maintains good lab practices for cleanliness. Helps
maintain good results. Check rotor fixing screw for
safety. Always inspect rotor for damage.
Maintains good lab practices for cleanliness. Helps
maintain good results. Always inspect buckets for
damage.
Helps to ensure that the rotor can be removed
for maintenance when needed. Does not affect
performance. Always inspect shaft for damage.
ERROR CODE – d (lid lock error)
Clearing Procedure:
• Turn the power off and wait 10 seconds
• Turn the power on
• Open and close lid
ERROR CODE – F (tachometer error)
Clearing Procedure:
• Wait for rotor stand still
• Turn the power off and wait 10 seconds
• Turn the power on
• Check to make sure rotor is installed properly
before spinning