Operation
1. Place the machine on a flat and stable surface. WARNING: Inspect power cord. If power cord frayed or damaged, do not use device
2. Inspect collection container, tubing, and inline filter for cracks or leaks. Replace all damaged accessories immediately.
3. Place the collection container into the collection container holder. Connect one end of the suction tubing to the patient port on the container lid.
4. As shown in the figure below, connect one end of the short intermediate tubing to the vacuum port on the collection container lid and connect the
other end of the short intermediate tubing to the "FLUID SIDE" of the inline filter. Connect one end of the long intermediate tubing to the inline
filter and the other end of the long intermediate tubing to the inlet of the machine.
5. Check all connections and make sure they are tight and secure.
6. Insert the plug into a wall (mains) outlet with a voltage matching the device's requirements. (See device label located on the underbelly of the device
for device's voltage requirements.) Turn the power switch on.
7. Completely cover the open end of the suction tubing, usually with thumb. Turn the vacuum adjustment knob (clockwise to increase vacuum pressure,
counter-clockwise to reduce) until the pressure indicated by the gauge equals the maximum pressure recommended by a trained healthcare
professional. WARNING: Never exceed this recommended maximum vacuum pressure.
8. Perform suction procedure in full accordance with the instructions of a trained healthcare professional.
9. Turn off the power switch and unplug the unit after each use.
inline filter
"FLUID SIDE"
short intermediate tubing
vacuum port
patient port
WARNING
• Prior to first use, carefully unpack the unit and check for transportation damage. In case of damage, do not use.
• Make sure the power switch is turned off before inserting and removing the plug.
• If the machine has been operating continuously for over 30 minutes, please turn off the power and allow the machine to cool off for at least 30 minutes
before turning it back on.
Cleaning, Storage, and Maintenance
1. Clean the product and all accessories thoroughly after each use. Prior to cleaning, make sure machine is turned off and unplugged.
2. Wipe machine clean with a damp cloth every few days to prevent accumulation of dust.
3. Do not clean with powder-type cleaning agents or soap. Do not clean machine in water.
4. Clean tubing after each use. Disconnect tubing from machine and clean by rinsing inside and outside of tubing with hot tap water. Air dry.
5. Wash collection container in hot tap water with dishwashing solution after each use. Air dry.
6. Keep the machine and relevant accessories dry. Avoid direct sun exposure during storage.
7. Before each use, inspect power cord for fraying/damage and all accessories (collection container, tubing, and inline filter) for cracks/leaks.
8. Change inline filter if collection container overflow occurs or every two months, whichever comes first.
WARNING
• Do not disassemble or modify machine. Maintenance should be performed by trained professional maintenance personnel only.
• This machine is an oil-less suction pump and does not require any lubrication. Do not add any lubricants to this machine.
• Do not clean machine by immersing in water. Wipe machine clean with damp cloth.
• To ensure proper hygiene and to prevent contamination and infection, each set of accessories should only be used by a single patient.
Troubleshooting
If you run into problems with the product, refer to the possible causes below to perform simple troubleshooting. If the problem cannot be solved, please
contact the distributor or manufacturer.
If machine does not start:
• Check to make sure unit is plugged into mains and that the power switch is in the "on" position.
• The aspirator automatically shuts off if it overheats. Allow machine to cool for 30 minutes before resuming operation.
• If the fibrous material inside the inline filter is wet, stained, or damaged, replace the filter.
If reduced performance:
• Check that mains provides the voltage and current levels required by this device.
• Check if any tubing is blocked.
• Make sure that all tubing is securely connected and that there are no leaks.
• Replace if collection container or tubing is damaged.
• If the fibrous material inside the inline filter is wet, stained, or damaged, replace the filter.
Copyright © 2019 ANEST IWATA SPARMAX Co., Ltd. All rights reserved.
long intermediate tubing
inlet
long intermediate tubing
short intermediate
tubing
Install inline filter between the collection container and machine,
with the word "FLUID SIDE" pointed towards the collection container.