EG&G ORTEC 402D Руководство по эксплуатации и обслуживанию - Страница 11

Просмотреть онлайн или скачать pdf Руководство по эксплуатации и обслуживанию для Источник питания EG&G ORTEC 402D. EG&G ORTEC 402D 16 страниц.

12
OUTPUT IMPEDANCE
<0.15n for the ±6-V outputs
and <0.3n for all other outputs at any frequency to 100
kHz.
TEMPERATURE COEFFICIENT
range of 0 to 60° C.
<0.02%/°C over a
TEMPERATURE PROTECTION
A thermal warning
switch will close when the supply temperature ap
proaches within 20° C of the safe operating value. A
thermal cutout switch disables the Power Supply when
the temperature exceeds the safe operating value.
NOISE AND RIPPLE <3 mV peak-to-peak for all six
outputs, as observed on a 50-MHz bandwidth oscillo
scope.
VOLTAGE ADJUSTMENTS ±2% minimum range, reset-
tablllty ±0.05% of Supply voltage.
CIRCUIT PROTECTION
Both sides of the Input line to
the power supply are fused. In addition, output current
foldback limiting to prevent damage to the Supply and
automatic recovery when the demand Is removed are
provided by electronic circuitry.
All six supplies are protected so that any one can be
shorted to any other one without resulting In permanent
damage.
Overvoltage protection Is provided on ±6 V so that these
outputs will not exceed 7.5 V maximum.
OUTPUT CONNECTOR
All power and control circuits
terminate In a connector, specified by TID-20893 (Rev 3),
which mates the Bin Interface connector, completing the
necessary control and Power Supply wiring.
DIMENSIONS
Conform to AEG drawing ND-515 and
paragraph L, page A-10, of TID-20893 (Rev 3).
RECOVERY TIME <100 fjs to return to within ±0.1% of
rated voltage for all six outputs for any change In Input
voltage and load current from 10% to 100% full load.
WEIGHT
Net 8.9 kg (20 lb); with 4001A(B) 13.6 kg (30 lb).
Shipping 12.3 kg (27 lb); with 4001A(B) 19.6 kg (43 lb).
3. INSTALLATION
The 402D Power Supply Is normally supplied factory-
connected to an EG&G ORTEG 4001A(B) Modular Sys
tem Bin. However, the Supply Is designed to TID-20893
(Rev) specifications and may be attached. In the space
provided, to any bin manufactured to TID-20893 (Rev)
specifications.
For attachment to other than an EG&G ORTEG 4001 A(B)
Bin, refer to the appropriate Instruction manual. The On-
Off switch and other controls necessary to operate the
supply are part of the Bin and are not furnished with the
Power Supply.
For attachment to the EG&G ORTEG 4001A(B) Bin the
following steps are advised:
1. Place the Bin on a table with the back part facing you.
Place the Power Supply In the proper mounting position,
leaving enough space between It and the Bin to attach
the Interface connector.
2. Mate the Interface connector, being careful to align
the polarizing pins. Fold and form all wiring close to the
connector edges to prevent any wires from being pinched
and producing a short circuit In succeeding steps.
3. Mount the Supply to the Bin by securely tightening
the four 10-32 screws, being careful not to pinch any
wires or to use undue force on any parts.
Unless otherwise specified, the 402D Is shipped with
transformer connections appropriate for operation on
nominal 117 or 230 V mains, according to the selection
made with the rear panel slide switch. To change the
connections for operation on 100 or 200 V power mains,
use the following steps:
1. Remove the top cover.
2. Move the wires from lug #3 of T1 to Its lug #2.
3. Move the wires from lug #6 of T1 to Its lug #5.
4. Replace the top cover.
5. Set the rear panel slide switch at 117 for 100 V opera
tion or at 230 for 200 V operation.