6.2 THAW FROZEN MILK TO COOL LIQUID STATE

Caution: the Penguin Nutritional Warmer is not recommended for
thawing less than 15 mL of breast milk in storage bags!
1. Scroll to choose the container type and size of interest.
2. Press Enter.
3. Only when ready to begin warming the feeding, place the feeding container into the inner
compartment (orange zipper) of the Therma-Liner and gently push it to the bottom. Press
firmly along the entire orange zipper to encompass the container as much as possible. Use care
when removing and re-inserting feeding containers to prevent damage to the interior water
compartment.
4. Place the prepared Therma-Liner in the well with the sealed opening at the top.
5. Press Enter*.
6. The device will begin warming, vibrating, and alternating screens between the profile selected
and elapsed time.
7. When the feeding has been warmed to a liquid cool state, the device will enter Thaw Complete
mode.
8. Please immediately remove the feeding** from the Therma-Liner and set the Therma-Liner
aside until next use during the shift.
9. Press On/Off to return the warmer to Standby mode.
*Note that if a wrong entry was made, you may press the Cancel selection button to go back a screen.
**If there are still some frozen particles in the feeding, let it sit until the remainder thaws, or rotate
the container gently to thaw the rest of the way. It is better to have a few remaining ice crystals than
to overshoot the refrigerated range. Any portion that is meant for immediate feeding may be placed
in a bottle or syringe for further warming. Feeding for the future should be placed immediately in the
refrigerator for up to 24 hours. After 24 hours, it should be discarded.

7. CLEANING

1. Disconnect Power Cord from the wall receptacle.
2. Prepare reagent wipes (e.g. CaviWipe™ or SaniWipe™) containing Isopropanol, Ethylene Glycol
Monobutyl Ether (2-Butoxyethanol), Diisobutylphenoxyethoxyethyldimethyl benzylammonium
chloride.
3. Using the prepared reagent wipe, clean and remove all surface debris from the device, including
the power cord.
4. Repeat as deemed necessary using a newly prepared reagent wipe each time.
5. After the last cleaning wipe is used, allow device to air dry for a minimum of 10 minutes or dry
with a soft cloth before returning the device to service.
11