Aktakom AM-1083 Kullanıcı Kılavuzu - Sayfa 2
Multimetre Aktakom AM-1083 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Aktakom AM-1083 5 sayfaları. Digital multimeter
I. Summary
This series product is a stable and battery-driven 3 1/2 digital
multimeter with high reliability. It adopts LCD screen with character
height of 28mm; with functions including unit sign/backlight and overload
protection, it is easy to use.
This series of device can be used to measure DCV, ACV, DC, AC,
resistance, capacitance, inductance, diode, triode, open circuit,
temperature and frequency, as well as hold peak value. As an
instrument with excellent performance it adopts double integral A/D
converter as its core.
II. For purchasers of this multimeter
1. Safety notice
This series of device is designed according to IEC1010 standard
(safety standard issued by International Electrotechnical Committee).
Please read these safety notices before using it.
1-1. Never input DC higher than 1000V or AC higher than 750V virtual
voltage as measuring voltage;
1-2. The voltage lower than 36V is safe. Please confirm the earthing,
connection and isolation of pencils to avoid electric shock when
measuring the voltage of DC higher than 36V or AC higher than 25V;
1-3. Pencils should be away from test points when switching function or
range;
1-4. Select correct function and range and beware of miss-operation.
You may still be careful to ensure the safety even the device has range
protection functions.
1-5. Never input current higher than 20A when measuring current;
1-6. Introduction for safety sign
"
" Dangerous voltage existing; "
isolation; "
" Shall refer to the user's manual; "
III. Unpacking inspection
Open the package to check if all parts and accessories are all right in
the box
1. Multimeter
1
2. Battery (9V)
3. Temperature probe (banana type)
4. Testing accessory
5. Pencil (20A)
6. User's manual
IV. Features
1. General features
1-1. Display mode: LCD;
1-2. Max display: 1999 (3 1/2) digits automatic polarity display;
1-3. Measurement method: double integral A/D conversion;
1-4. Sampling rate: about 3 time per second;
1-5. Over-range display: the highest digit displays "OL" or "-OL"
1-6. Low voltage display:"
1-7. Working environment: (0~40) , relative humidity: <80%;
1-8. Power supply: one 9V battery (6F22 or equivalent model);
1-9. Dimension: 189mm * 97mm * 35mm (length*width*height);
1-10. Weight: about 400g (including 9V battery);
2. Technical features
www.tmatlantic.com
" Earthing; "
" Low battery.
1
1 pair
1
1 pair
1
";
2-1. Accuracy: ±(a% × reading data + character number), ensure in
accuracy environment temperature:
(23±5) , relative humidity <75%, and the calibration warranty period
lasts for one year from the date of leaving the factory.
2-2. Performance (▲ indicates the function is available for this model)
Function
DCV
ACV
DCA
ACA
Resistance
Diode/ Open circuit
Triode
Capacitance
Temperature
Frequency
Inductance
Auto power off
Backlight
Peak value holding
2-3. Technical index
2-3-1. DCV
Range
200mV
2V
20V
200V
1000V
Input impedance: All range is10MΩ;
Over load protection: 250V DC or 205 V AC peak value for range
" Double
200mV; 1000V DC or 1000V AC peak value for other.
2-3-2. ACV
Range
200mV
2V
20V
200V
750V
Input impedance: All range is10MΩ;
Over load protection: 250V DC or 205 V AC peak value for range
200mV; 1000V DC or 1000V AC peak value for other.
Frequency response: 40~400Hz for range under 200V; 40~100Hz for
range 750V;
Display: Sine wave virtual value (AVG value response).
2-3-3. DCA
Range
2mA
20mA
200mA
20A
MAX measurement voltage drop: 200mV;
Max input current: 20A (no more than 10 seconds);
Over load protection: 0.2A / 250V self-healing fuse for ranges under
200mA, 12A / 250V quick acting fuse for range 20A.
2-3-4. ACA
Range
2mA
Ω
hFE
C
f
L
Accuracy
±(0.5%+3)
±(1.0%+5)
Accuracy
±(1.2%+3)
±(0.8%+5)
750V
Accuracy
±(0.8%+3)
±(1.2%+4)
±(2.0%+5)
Accuracy
±(1.0%+5)
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
Resolution
100uV
1mV
10mV
100mV
1V
Resolution
100uV
1mV
10mV
100mV
1V
Resolution
1uA
10uA
100uA
10mA
Resolution
1uA