3. Connect the black test lead to COM jack and the
red test lead to V: jack.
4. Touch the probes to the test points. In ohms, the
value indicated in the display is the measured
value of resistance. In continuity test, the beeper
sounds continuously, if the resistance is less than
100:.
5. When using 400M: Range; The 400M: Range
has a fixed 20+2-count offset in the reading.
When the test leads are shorted together in this
range, the meter will display 02.0. The residual
reading must be subtracted from the reading
obtained in step 4 when this range is used. For
example, when measuring 100M: on the 400m:
range, the display will read 102.0, from which the
20 residual is subtracted to obtain the actual
resistance of 100.0M:.