Data String Format Table
Prefix (3 bytes)
S = Sensor number R = Range number
P(S)(R)
R = Range number
mA(R)
S = Sensor number
P(S)T
Amb
x = Z for zero -or- x = S for span
BxC
C = ADC channel number
= Delimiter
,
n/a = no delimiter is present for marked location in that string
= Terminator: >
or
or
<
?
*
> = good battery
< = low battery
? = expired battery (readings are not valid)
Model Ranges
The standard Model 33 range serial numbers 2262*-xxxxxx are:
LP Sensor
HP Sensor
(Eight Ranges)
(Three Ranges)
P11= inH
O
P21= bar
P21= kg/cm
2
P12= mbar
P22= kPa
P22= bar
P13= kg/cm
P23= PSI
P23= Mpa
2
P14= mmHg
P24= kPa
P15=mmH
O
P25= PSI
2
P16= kPa
P17= inHg
P18= PSI
,
Data
,
8 bytes of direct ADC output
,
8 bytes of direct ADC output
8 bytes of direct ADC output of sensor
,
temperature signal
8 bytes of direct ADC output of sensor
,
temperature signal
27 bytes of ASCII spaces.
n/a
This is a background calibration of the
analog to digital converter - no data.
(Five Ranges)
2
,
Data
,
Data
,
8 bytes of displayed data
,
8 bytes of internal offset/tare value
,
8 bytes of displayed data
n/a
18 bytes of ASCII spaces
n/a
18 bytes of ASCII spaces
*
n/a
18 bytes of ASCII spaces
*
*
3
Communication
,
*
*
30 Series Programming Instruction Manual