Doran DS8150 Kullanıcı Kılavuzu - Sayfa 10

Terazi Doran DS8150 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Doran DS8150 11 sayfaları. Chair scale

Doran DS8150 Kullanıcı Kılavuzu

Section 5. Troubleshooting

General Problem Resolution

Weight reading will not repeat or
scale does not return to zero when
weight is removed.
Scale overloads before scale
capacity is reached.
Scale will not come to zero when the
Zero button is pressed.
Weight readings are not correct.
Scale drifts off zero.
Scale shuts itself off or will not turn
on.
8
Problem
What to Do or Check
Person or object other than that being
weighed is touching the scale.
Verify that there is nothing caught in or
interfering with the seat.
Verify scale calibration is correct. If
problem persists, recalibrate the scale.
Make sure that the scale is stabilizing
(Motion annunciator is off). After pressing
the Zero button, the scale should zero as
soon as it becomes stable.
Verify the scale calibration with an
accurate test weight. If the readings are
not correct, recalibrate.
Be sure the chair is on a flat surface and
all four wheels are touching the floor.
Check for air currents and/or vibration
around the scale.
Press the Zero button to turn on the
indicator.
Check that new batteries are installed
properly in the battery tray.
If using an AC adapter, confirm that it is
firmly plugged into a functioning wall
outlet.
Verify the connector at the rear of the
indicator is properly installed.