P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement
Technical Specification!
!
Please refer to the MicroBio Manual P0001W003 for full detailed specification of the
MicroBio MB2.!
!
The MB2-RSH model is capable of driving a remote sampling fan up to a distance of 20m
from the control unit.!
!
Installation!
!
The first task is to run the connecting cable from where the sampling head is to be
positioned to where the control unit is to be located.!
!
Two cable options can be supplied. One is with an extension cable with connectors at
each end, the other with one of the connectors not fitted. By not terminating one of the
connectors enables the cable to be passed through glands or bulkheads as required by
the end user. Once the cable is routed, the unterminated end, located where the control
unit is, must have the supplied locking connector fitted. Note that the pre-fitted connector
on the extension cable, then end labelled "RSH", must be where the remote sampling
head is located. !
!
Fitting the Connector for Unterminated Cables!
!
The following tools are required to fit the connector:!
!
1 x Soldering iron and solder (lead free)!
1 x 2.5mm slot screwdriver!
!
The connector is assembled in the order as shown in the photographs below.!
!
1. Dismantle the connector by first removing the screw on the side of the connector.

!