Follett Edge Series Kurulum Kılavuzu - Sayfa 4
Buzdolabı Follett Edge Series için çevrimiçi göz atın veya pdf Kurulum Kılavuzu indirin. Follett Edge Series 17 sayfaları. Laboratory freezer
Ayrıca Follett Edge Series için: Kurulum Kılavuzu (17 sayfalar)
Dimensions
Width (up to hinge cover)
Height (up to hinge cover)
Depth (up to Solid Door)
Door Height
Electrical Specifications
§ 115 V, 60 Hz, 1 phase (Main supply voltage fluctuations not to exceed ±10% of the nominal voltage (120 V)
Refrigeration Specifications
Refrigerant
R600a
Refrigeration System
The refrigeration system is designed to give many years of trouble-free service. Except for routine cleaning of the
air-cooled condenser and related parts, the refrigeration system requires no service or maintenance. The system
uses a capillary tube and is critically charged. The connection of refrigeration service hoses to the fittings will almost
invariably result in a significant change in the system charge. This change can adversely affect the performance
of your refrigerator. Therefore, Follett recommends that if hoses are ever connected to the refrigeration system
for service, the refrigerant should be recovered, the system evacuated, and recharged by weighing in the correct
refrigerant charge.
Use of a combustible gas meter for detecting HC refrigerants needs to be with in 1 meter of working area. Service
technician should also have a safety placard for the work area stating "Flammable Gas no smoking or open
flames".
Installation Specifications
§ Altitude up to 6561.68 ft (2000 m)
§ Ambient temperature 68 F to 75 F (20 C to 24 C) for maximum efficiency
§ Maximum relative humidity 75% for temperature up to 24 C decreasing linearly to 50%
§ Ambient temperature must not exceed 100 F (39 C)
§ The front louvered panel must be kept free of any cabinet trim or obstructions to ensure proper ventilation of
the refrigeration system - allow at least 12" (30.48 cm) of clearance in front of panel
CAUTION!
§ Equipment must be wired according to local and national electrical codes.
§ Always disconnect power before servicing refrigerator.
4
REFNAT5-SD, REFNAT5-GD
23.84" (60.6 cm)
34.16" (86.7 cm)
23.66" (60.0 cm)
SD: 30.63" (77.7 cm), GD: 30.75" (78.1 cm)
Charge Size
52 g (0.114 lb).