O
PERATION
The instructions for operating the temperature controller are given in a separate manual.
If the furnace is fitted with a time switch, see also the supplementary manual MS03.
3.1
Operating Cycle
If the furnace has an auxiliary heater, see also section 3.2.
The furnace is fitted with a combined Supply light and Instrument switch. The light is on
whenever the furnace is connected to the supply. The switch cuts off power to the control circuit.
There is also a Heater switch which can be used to disconnect power to the elements.
Connect the furnace to the electrical supply. The Supply light should glow.
Operate the instrument switch to activate the temperature controller; the O position is off, the
I position on. The controller becomes illuminated and goes through a short test cycle. Adjust the
temperature controller (see controller manual).
Overtemperature option only. If the overtemperature controller has not yet been set as required,
set it and activate it according to the instructions in the appropriate manual.
Switch on the Heater switch, located on the instrument panel.
Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light(s) glow
steadily at first and then flash as the furnace approaches the desired temperature or a program
setpoint.
Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light(s) glow
steadily at first and then flash as the furnace approaches the desired temperature or a program
setpoint.
Overtemperature option only. If the overtemperature trip operates then an indicator in the
overtemperature controller flashes, and the heating elements are isolated. Find and correct the
cause before resetting the overtemperature controller according the instructions supplied.
To switch off power to the heating elements, use the Heater Switch .
To switch the furnace off, set the Instrument switch to off. If the furnace is to be left off, isolate it
the electrical supply.
3.2
Auxiliary heater (if fitted)
The auxiliary heater is controlled by an independent temperature controller, which is affected by
the same instrument switch as the main controller. Note that a time switch (if fitted) may not
switch off the auxiliary heater, depending on other options ordered.
Set the auxiliary heater's temperature controller after setting the main controller, as given in
section.
Auxiliary heater controller navigation diagram:
Home
List
3.3
20.0
measured temperature;
*C
use
setpoint
OP
output power,
read only
MF10
*C
/
to access
for factory access to
lists and parameters
not available to the
operator
CFM
Access
List
ACCS
codE
5