5.1 Calibration
5.1 Calibration
Main menu
Calibrate
.
.
.
.
Basic settings
Calibration procedure
1)
Switch the controller OFF and select manual temperature compensation. Enter
the temperature of the calibration solutions.
2)
Immerse the electrode in one of the calibration solutions. Wait until the
measured value is stable, then calibrate by pressing keys and - while still
applying pressure - additionally key
now displayed as measured value.
3)
Rinse the electrode and repeat step 2 with the second calibration solution.
4)
Check the slope and sensor zero-point, then put the electrode back into the
armature. Select automatic temperature compensation and switch ON the
controller.
NOTE
Buffers
At works the following buffers are stored:
Calibration solutions:
Inner buffer of the electrode:
If you want to use other calibration solutions, adjust the values - it does not matter if
you start with the higher or the lower pH.
If you are using electrodes with a special inner buffer, adjust the pH value of the sensor
buffer, since this value is used as zero for the calculation of the measured values. You
will find the pH of the sensor buffer printed on the electrode.
20
Calibrate
Meas. val. 6.99pH
Cal. press &
Slope
058 mV/pH
Sensor zero
001 mV
Basic settings
.
Buffers used
The slope should be close to 59 mV, the zero-point close to 0 mV. The
slope decreases and the zero error increases with time. When either
value exceeds certain limits, the instrument displays an error
message indicating that the electrode has to be replaced.
buffer solutions pH 4 and pH 7
pH 7
K 100 PR & K 100 W PR
The calibration solutions stored as
„buffers used" are recognised
automatically. It does not matter whether
you start with buffer 1 or 2.
Buffers used
Buffer 1
7.00 pH
Buffer 2
4.00 pH
Sensor buffer
7.00 pH
. The pH of the calibration solution is