Mach 380 / Mach M3 Stand model
Caution:
Always draw plug from mains before doing any work at the lamp, stand or power supply.
The electrical connection between stand tube, spring arm and lamp is ensured by sliding con-
tacts.
Connect the mains supply.
Check the function of the lamp.
15
59520001
6. Electrical connection
Threre can be connected three additional de-
vices to the plug strip of the stand.
Attention: We recommend to consult an
16
electrician before connecting additional
devices in order to avoid electrical overload
of the stand.
Leakage current
When several devices are connected to one
main plug, the sum of the leakage currents
may exceed the maximum limit.
16
Additional devices
In case the device is used in the patient
area together with other devices, these de-
vices must be supplied by a isolation trans-
former, or connected to the earth protection
system, if the additional devices do not
comply with standards IEC 601-1 and IEC
601-1-1.
For safety reasons the safety plugs are cov-
ered with the cover 15.
For connecting additional devices proceed as
follows:
• Turn off the lamp.
• Loosen the four screws 16 with an Allen key
• Connect the additional devices.
• Fix the cover 15 with the four screws 16.
Edition 10
!
DANGER
width 5 and remove the cover 15.
11.08.2004 / Bak Page 9/14