Advanced Instruments OsmoTECH PRO Hızlı Başlangıç Kılavuzu - Sayfa 3

Tıbbi Ekipman Advanced Instruments OsmoTECH PRO için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Advanced Instruments OsmoTECH PRO 5 sayfaları. Sample probe motor
Ayrıca Advanced Instruments OsmoTECH PRO için: Hızlı Başlangıç Kılavuzu (4 sayfalar)

Advanced Instruments OsmoTECH PRO Hızlı Başlangıç Kılavuzu
Removing the sample cooling assembly
1.
Remove the four screws and washers [F] that
secure the cooling fan, the finger guard, and the
air‑flow diverter to the cooling assembly's heatsink.
Do not remove the cooling fan wire leads from the
cable clamps.
NOTE:
Remember/record the position and
orientation of the cooling fan before
removing it from the chassis.
2.
Disconnect the sample well drain tubing [G] from
the drain tube [H].
G
3.
Remove the four socket head cap screws and
washers [I] that are used to secure the sample
cooling assembly to the sample well mounting
block.
I
One of the socket head cap screws also secures a
cable clip.
4.
5.
6.
H
I
Replacement Procedure
OsmoTECH® PRO Block Probe
Remove the block probe cable, the thermo‑electric
cable, and the solenoid cable from the two cable
clamps [J].
J
Detach the cable clip [K] from the thermo‑electric
wire leads.
K
Carefully lift the sample cooling assembly out of the
osmometer module assembly.
J
Page 3 222123PM Rev 0