Symbols
Turns on or off the device
Set the clock / Does not store the
current data
Direction guide to install battery
Direct current
SN
Serial number
Date of manufacture
Type BF: Device, cuff and tubing
are designed to provide special
protection against electric shocks.
Appears when measurement is in
progress and flashes when the
pulse is detected.
Previous measurement stored in
MEMORY
Appears when the battery voltage is
too low for the device to work
properly.
Appears if the systolic and diastolic
measurements are within 10 mmHg
of each other.
Appears if the pressure value is
unstable due to movement during
the measurement.
Appears if the pressure value did
not increase during pump up.
System error
Appears if pulses are not detected
correctly.
Attention symbol
Symbols
Function / Meaning
4
Treatment
___________
___________
___________
___________
___________
___________
___________
Measurement is in progress --
remain as still as possible.
___________
Replace all batteries with new
ones.
Verify that the air hose is
properly connected and that
the exhaust rate is between 2
and 5 mmHg. Check for air
leakage.
Try the measurement again.
Remain very still during the
measurement.
Verify that the air hose is
properly connected and that
the exhaust rate is between 2
and 5 mmHg. Check for air
leakage.
Reload batteries.
Attach the cuff properly and try
the measurement again.
See instruction for use