ANDalyze AND1100 Kullanıcı Kılavuzu - Sayfa 10
Ölçüm Cihazları ANDalyze AND1100 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. ANDalyze AND1100 16 sayfaları. Fluorimeter for water testing
AND1100 Fluorimeter for Water Testing
Scale: Slope of the line that correlates the rate of fluorescence
increase vs. concentration. This value is determined by the 3‐
point calibration. Some sensors have a default scale that is
automatically calculated from other sensor's scales. In this
case, "Auto" will be indicated in the default scale field. A new
3‐point calibration will supersede the default settings.
Offset: Intercept of the line that correlates the rate of
fluorescence increase vs. concentration. Also automatically
determined by the 3‐point calibration.
3‐Point Lab Calibration
Using the left/right keys in the
Advanced menu, Select the
Calibrate option.
When all (3) solutions have been
tested, the calibration results
screen will be displayed. Click
Apply to save the results. The
settings will then be viewable in
the Advanced screen.
4.7 Firmware Update
ANDalyze is continually adding new features and options to the AND1100 through new firmware updates. To
update your AND1100 firmware, please visit andalyze.com and log in to register your device. Once logged in,
you will see the option to update your device. Follow the instructions.
ANDalyze Inc. 800 Boylston Street, Prudential Tower, Floor 16, Boston, MA 02199 USA Tel. +1 857.453.6740 www.andalyze.com
Note: This is for laboratory
calibration ONLY and
requires (3) standardized
solutions from
manufacturer or prepared
by a laboratory technician.
Delay: Points from the start of the test which
are not used for calculations. This is used
when there is a delayed reaction in certain
sensors.
R2: R‐Squared is a statistical term indicating
how good one term is at predicting another. In
our case, this is derived from the best fit line
created by the 3‐point lab calibration. Values
closer to 1 are better.
Factory Reset
The Reset option will restore the setting of that
specific sensor to factory settings.
Using the left/right keys in the Advanced menu,
Select the Reset option.
Select Reset on the Warning screen to complete
the process.
Page 10 of 16
July 1, 2013