ATI Technologies B12-69 O & M El Kitabı - Sayfa 6
Verici ATI Technologies B12-69 için çevrimiçi göz atın veya pdf O & M El Kitabı indirin. ATI Technologies B12-69 20 sayfaları. 2-wire wet oxidant gas transmitter
Model B12-69 Wet Oxidant Gas Transmitter
Type: Wet Oxidant Gases such as Cl
Range: 0-10 PPM Min., 0-100 PPM Max.
Accuracy: ±0.2 PPM at constant temperature
Electronic Repeatability: ±1% of span
Electronic Linearity: ±0.5% of span
Output: Loop-powered 4-20 mA, 675 ohms maximum at 24 VDC
Power: 12-28 VDC
Display:
None. Optional LCD available.
Enclosure:
NEMA 4X polystyrene
Explosion-proof cast iron optional, Class 1, Div. 1, Groups B, C, & D.
Controls: Zero and span potentiometers mounted on transmitter circuit board
Operating Temperature: -5º to +55º C
Sensor: Model 69 Wet Oxidant Gas Sensor
Weight: 0.25 Lbs (0.12 Kg.)
O & M Manual
Rev-G, 7/15
SPECIFICATIONS
, Br
, and ClO
2
2
2
- 6 -