MESSAGE
WHAT IT MEANS
Use a used test strip.
Error massage indicates
that you may remove the
strip after applying blood
to the absorbent hole.
Problem with the meter.
Environmental tem-
perature is lower than
10 (50 ) or higher
than 40 (104 )
Display remains
1. Battery may be
blank after the
installed improperly.
test strip has
2. Battery power is
been inserted
too low for use.
into the Meter.
3. Too much time has
passed between
inserting the Test
Strip and perfor-
ming the test.
31
ACTION
Please use a new strip
and re-test.
Try again with a new
test strip.
Re-test with a new test
strip. If the problem is
still unsolved, please
call the Customer Care
Line for help.
Please test at an envi-
ronmental temperature
1.Install the Battery
correctly.
2.Replace with new
Battery.
3.Reinsert the Test
Strip into the
Meter.
4. Test Strip has not
been fully inserted
into the Meter.
Test results
1. Not enough sample
are incon
in the Test Strip.
sistent or
2. Test Strip or Control
Control
Solution has expired.
Solution test
3. Test Strip has been
results are
damaged due to
not within
heat or humidity so
the specified
that sample cannot
range.
be applied or the
speed of applying
sample is too slow.
4. System is not per-
forming due to the
environment being
above or below
room temperature.
The Meter co-
Test Strip has not
untdown did
been inserted
not start.
correctly.
4. Reinsert the Test
Strip into the Meter.
1. Redo test with new
test strip and make
sure that enough
sample has been
added.
2. Redo test with new
test strip.
3. Perform a Control
Solution test using a
new Test Strip. If
results are still out of
range, replace with
new vial of Test Strips.
4. Bring System to a
room temperature
environment and
wait approximately 30
minutes before per-
forming a new test.
Use a new Test Strip
and redo the test.
32