Alert J5 Kullanım Kılavuzu - Sayfa 10
Test Ekipmanları Alert J5 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. Alert J5 20 sayfaları.
Activating
ALERT J5 is shipped deactivated after it is calibrated
and passes inspection at the factory. This allows the
device to be stored without losing any time from the
calibration period. You have to activate the device to start
using it. Once activated, the date for the next calibration
increments by 365 days from the date of activation.
A Not Activated message when you initially power on
the device indicates that it is not activated and no test
can be performed until it is activated.
NOTE: This feature is displayed on the Main Menu, only
if your ALERT J5 has been previously calibrated. It is not
possible for you to deactivate the device.
To activate:
1. From the Not Activated screen, press the left button
to activate it. The date the device was deactivated is
shown.
2. Press the right button to activate the device. The
calibration due date is now 365 days from the current
date.
Battery
The ALERT J5 requires two AA Alkaline batteries to
operate. To insert, remove the battery compartment rear
cover (press the tab to release).
p. 4/13
ALERT
J5
TM