Classic infusion lines
Essential precaution
= the infusion bag, the entire drip chamber and the infusion line must be completely purged of any
1. Close the infusion line roller clamp
2. Insert the spike into the infusion bag like for a classic
infusion
3.
Place the bag in the horizontally arranged Droper Field
4. Close the Droper Field
5. Install the Droper Field on its back side by
placing it vertically
6.
Open the infusion line roller
clamp
7. Purge all air contained in the entire system (bag,
entire drip chamber and line).
8. Close the infusion line roller clamp
9.
10. Connect the infusion line to the patient and re-open the roller clamp
Flow regulation : through a flow regulator if applicable
or simply by estimating the flow by the relative position
of the roller clamp.
:
carry out the complete backwards purge
contained air.
4
6
8
Return the Droper Field to the horizontal
position, even if it can work in any position
(vertical, lateral, inverted, hanged, etc.)
Instructions for use - Version 2
1
2
5
7
9
10
3
8/15