Droplet SP2 Manuel - Sayfa 4
Ölçüm Cihazları Droplet SP2 için çevrimiçi göz atın veya pdf Manuel indirin. Droplet SP2 13 sayfaları. Beam-scan camera and software
Introduction
Optimum performance of the SP2 is obtained when the YAG laser beam is operating in the
TEM
mode. This is defined by a single laser beam spot which has a Gaussian profile. Higher-
oo
order modes will have multiple spots, which may not affect the measurement of black carbon
incandescence but will almost certainly limit the utility of the scattering data collected. In cases
where very high-order modes are present, the incandescence data can also be compromised.
The SP2 beam scan camera and software is designed to facilitate observing and setting the
mode of the YAG laser in the SP2. The system consists of a Mightex CCD camera which has been
fitted with a high Optical Density (OD) filter. A mounting bracket allows the camera to be
mounted in the position normally occupied by the YAG power monitor on the SP2. The light-
intensity data from the camera is recorded and processed by a software package made by
Visulux, called vBeam. In the Windows 7 32-bit version, the vBeam software has code in it to
read only a camera with a specific serial number. If multiple units of the camera and software
are purchased, the installed software must be correlated with a given camera. With Windows 7
64-bit version and newer, the drivers should enable the software to find the correct serial
number of your camera automatically. In both versions of Windows 7, it is necessary to load the
software thru both USB ports so that the camera will be found by the system. If the software is
loaded into one port and the camera is plugged into the other USB port, the camera will not be
found.
Installation
For installation of the software, follow the instructions given in the Visulux manual. This manual
will discuss the installation of both the driver for the camera and the Visulux executable
software. In the setup of the Visulux software, one minor change from the default values is
recommended. The software will be installed in the folder Program Files/Visulux. Navigate to
the Visulux folder and open the file mightex.prop using Notepad. The following shows the
contents of this file. The default for the exposure time is 10000. This will result in saturation of
the camera response, and it is recommended that this be set at 5000 for a starting value. This
can be adjusted later as needed.
4
DOC-0175 Rev D
© 2017 DROPLET MEASUREMENT TECHNOLOGIES