Ankom A2000I Servis Prosedürü

Ölçüm Cihazları Ankom A2000I için çevrimiçi göz atın veya pdf Servis Prosedürü indirin. Ankom A2000I 2 sayfaları. Motor assembly replacement
Ayrıca Ankom A2000I için: Servis Prosedürü (3 sayfalar), Servis Prosedürü (3 sayfalar), Değiştirme Kılavuzu (2 sayfalar)

Ankom A2000I Servis Prosedürü
Port B
Valve Assembly
Flush analyzer ports A and B.
1.
a. Turn the instrument power switch to the ON position. The
Display will light.
b. Verify that the drain hose is securely in the drain.
Note: The water used to flush ports A and B goes directly
out the instrument's drain.
c. Press the arrow keys on the Keypad until you see "Select
Flush" on the Display.
d. Attach the Amylase Dispenser to port B on the right side
of the instrument.
e. Press ENTER on the Keypad. This will open the port A,
port B, and drain valves, and run the agitator motor.
f. Fill the dispenser with hot water.
g. Attach the Amylase Dispenser to port A on the left side of
the instrument.
h. Fill the dispenser with hot water.
Note: When ENTER is pressed on the Keypad, the
agitator motor will run for 2 minutes and then turn off. If
you complete the flush before 2 minutes, you can press
ENTER a second time to turn off the agitator.
ANKOM Technology  2052 O'Neil Road  Email: [email protected]  Phone: 315-986-8090  Fax: 315-986-8091
A and B Valve Cleaning
A2000/A2000I
Instrument Rear View
Service Procedure
Revised: 08/05/2014
Port A
Valve Assembly
Display
Amylase Dispenser
004
Keypad
pg. 1 of 2