Druck DPI 841 Kullanıcı Kılavuzu - Sayfa 7
Test Ekipmanları Druck DPI 841 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Druck DPI 841 14 sayfaları. Frequency calibrator and frequency loop calibrator
Example 2) Set up values for x:y Scaling = 0 to 100%.
Settings
Settings
1
x:y
▲
▼
Sk1 = Edit
= Flow scaling (mA, pressure only)
Table 5: (Part of table) Menu options - Settings (Output)
Options
Description
Hz
... Units Frequency Units. To select one of these units:
Hz: Range < 1000Hz
Pulses/minute (cpm)
Amplitude
To set the amplitude and the mode of the output
signal. Amplitude = 5V (Default).
Bipolar mode [✓]/[ ]: Set this option to make the
signal pass thru zero.
= Unipolar
Waveform
To set the waveform for the output signal:
,
,
To select and set up a value for the "Nudge" output.
Example: 0.010 kHz increments.
Additional data: Select Settings (■ ■)
To select and set up values for the "Span Check"
output. Example output cycle:
Additional data (Table 6): Select Settings (■ ■)
To select and set up values for the "% Step" output.
% Step
Example output cycle:
Additional data (Table 6): Select Settings (■ ■)
K395 Issue 3
2
Edit
x:y
▲
■ ■
▼
■ ■
kHz: Range 1 ... 50kHz
Pulses/hour (cph)
= Bipolar
= Sine, square, or triangle
This cycle repeats
automatically.
Auto Repeat - Optional
Table 5: (Part of table) Menu options - Settings (Output)
Options
3
4
Hz
... Step
Options
Pulses
Table 6: Additional data for Settings (Output):
Item
Span Check
Low (0%)
High (100%)
Dwell (d)
% Step
Step Size (s) ...
%
Defined Step Low (0%), High (100%), Dwell (d): As above.
Step Size (s)
Ramp
Travel (t)
Auto Repeat
Description
To select and set up values for the "Defined Step"
output. Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings (■ ■)
To select and set up values for the "Ramp" output.
Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings (■ ■)
Description
Counts: To set the total number of output pulses.
Rate: To set the frequency of the output.
Amplitude: To set the amplitude of the output
signal. Amplitude = 5V (Default).
Bipolar mode [✓]/[ ]: Set this option to make the
signal pass thru zero.
= Unipolar
Value
Set the 0% value.
Set the 100% value.
Set the period (Hours:Minutes:Seconds) between
each change in value.
Low (0%), High (100%), Dwell (d): As above.
Set the change in value for each step as a
percentage of the full-scale range (High - Low).
Set the change in value for each step as a
frequency value.
Low (0%), High (100%), Dwell (d): As above.
Set the period (Hours:Minutes:Seconds) to go from
the Low (0%) value to the High (100%) value.
If applicable, select this item to repeat a cycle
continuously.
= Bipolar
5